FUTURE IS OUR FIRST PRIORITY
NATALLY Radiant Infant Warmer
Creating a safe, comfortable, and nurturing environment for newborns is essential, both in delivery rooms and during subsequent intensive care. This includes ensuring optimal temperature control, minimizing noise and light disturbances, and providing state-of-the-art medical equipment designed to meet the delicate needs of infants.
The goal is to create a space where newborns can feel secure, promoting their health and development while also offering parents peace of mind during these critical early moments.

Key Features
- Preheating mode, manual mode and body temperature mode
- 90° rotation of the emitter to both sides
- Electrical bed tilting
- Positionable 9” colour LCD monitor
- Vital sings monitoring by Masimo
- Integrated weighing scale
- Height-adjustable chassis with two-side control
- Phototherapeutic module
- Wide selection of accessories
Technical Specifications
TECHNICAL SPECIFICATIONS: Radiant Warmer Natally RW-610 | |
---|---|
Unit type identification | RW-610 |
Power voltage | 110 / 120 /127 / 230 V, 60,50 Hz |
Power input | 700 VA |
Dimensions |
Height without height adjustable chassis: 1 895 mm Height with height adjustable chassis: 1 835 - 2 035 mm Length: 1 020 mm Width: 580 mm Mattress: 730 x 485 x 40 mm |
Sound pressure level (audible alarm signal) | 66 – 72 dBA (measurement radius of 1 m) |
Maximum input via auxiliary sockets | 100 W |
Weight |
Natally with fixed height chassis: 92 kg Natally with height adjustable chassis: 102 kg |
Rotating of the emitter | 180° (by 90° anti- and clockwise) |
Electric bed positioning | 12° (Trendelenburg, anti-Trendelenburg) |
TEMPERATURE MODE | |
Manual mode (adjustment range) | 1 – 100% heating performance, 1% steps |
Range of displayed performance values | 0 - 100 % |
Body temperature mode (adjustment range) | 34,0 °C - 37,0 °C, in steps of 0,1 °C 37,1 °C - 38,0 °C, in steps of 0,1 °C |
Body temperature adjustment over 37.0 °C | Only upon special operator’s intervention after the >37 °C key is pressed |
Range of displayed temperature values | 10,0 °C - 45,0 °C |
Preheating mode | Allows immediate care of the newborn child after delivery |
APGAR SCORE MODE | |
Range of the measured time | max. 99 min |
Range of displayed values | 0 - 99 min |
Accuracy of time measurement | ±0,1 s |
Resolution | After 1 s |
WEIGHING MODE | |
Scales range | max. 7 kg |
Range of displayed weight values | 0 - 7 kg |
Scale accuracy |
±2 g ±1 digit (from 0 kg to 1 kg) ±5 g ±1 digit (from 1 kg to 3 kg) ±10 g ±1 digit (from 3 kg to 7 kg) |
Resolution | 1 g |
Records | Weight trends, weight increments |